Home GC112A Gas Chromatograph

GC112A Gas Chromatograph




GC112A Gas Chromatograph


Features


1. The gas chromatograph offers high accuracy oftemperature control (optimum: ± 0.05°C), high reliability andanti-interference.
2. Control and operating parameters, including for detectors,can be set up by the keyboard. Settings include elf-diagnosis system,power-fail protection, file storage and load, and limited temperature. The gaschromatograph uses accurate display of setting and actual value of temperaturecontrol of each path, bridge current of TCD, sensitivity of FID, retention andanalysis time.
3. It offers four paths of independent controlsections, including column oven, sample injector, detector and thermalconductive pond.
4. It has 5-rampprogrammed temperature rising control and auto-drop of temperature of columnoven to a fast cooling.
5. Dual FID providedwith options of GC112 A-TCD, GC 112 A-FPD and GC112A-NPD, gas sampling valve,transformation oven, absorber and pyrolysis absorber are available forselection. Two detectors can be installed at the same time with highsensitivity. It takes a short time to be stabilized, it is easy to clean andinstall nozzle.
6. A packed column and capillary column (0.53 large diameter orsmall diameter) can be easily installed including a complete set of split samplingconnector and tailing connector to be installed free of charge. An optional 0.53mm large diametercapillary direct sampler is available for choice.
7. Analysis can be conducted in packed or capillary column with multiplesampling methods.
8. A packed column can finish on-column sampling, instantaneousvaporized sampling and gas sampling.
9. The gas chromatograph usescapillary column, which can support split sampling and 0.53mm large diameter capillarydirect sampling.
10. Construction of dual gas path, dual sampleinjector and dual packed column system can effectively avoid baselineundulation and drift in real time and facilitate programmed temperature risinganalysis. Accurate and scaled type gas path control valves will providerepeatability and stability of the gas chromatograph.
11. The gas chromatograph supports an auto ON/OFFtemperature control system, permanent data storage, file storage and call andover-temperature protection.
12. The large capacity column oven (300x280x270mm) can hold capillarycolumn and dual packed column. Built-in heating filament structure is in favorof quartz capillary analysis.
13. A complete set of accessories includingpurifier, pressure-reducing valve, gas path tube, wrench and tools, injectionneedle and connectors of all kinds are provided free of charge and are ready toconnect with the supply-air system prepared by users.

TechnicalSpecifications

GC112A Gas Chromatograph

TemperatureControl


Control range

7°C - 400°C above ambient. (increment of 1°C)

Control objects

Column oven, detector, thermal conductive pond and sample injector

Programmed temperature rising

5 ramps

Rate of programmed temperature rising

0.1°C -40°C (increment of 0.1°C)

Thermostatic time

0-655 min (increment of 1 min)



FID


Sensitivity (minimum testability)

M£5´10-11 g/s (sample: C16)

Optimum testing result

M£1´10-11 g/s

Noise

£5´10-14 A

Drift

£6´10-13 A/h

Linear range

³106



TCD


Sensitivity

³2000mV.ml/mg

Noise

£20mV

Drift

£30mV

Linear range

³104



FPD


Sensitivity (minimum testability)

£2´10-11 g (P)/s (sample P in the parathion methyl)

Optimum testing result: £3.8´10-12 g (P)/s.

Sensitivity (minimum testability)

£1´10-10 g (S)/s (sample: S in the parathion methyl)

Optimum testing result: £2.7´10-11 g (S)/s

Linear range

³103 (sample P parathion)

³102 (sample S parathion)

Stability

£30mV/h

Noise

£10mV

Starting time

Not greater than 2 hours



NPD


Sensitivity (minimum testability)

Dt (N)£5pg (N)/s

Optimum testing result: 0.5ppm

Sensitivity (minimum testability)

Dt(P) £0.5pg (P)/s

Optimum testing result: 0.2ppm

Drift

£8´10-13 A/h

Noise

£5´10-13 A

Linear range

³103

Max application temperature

360°C



Optional Accessories


1. TCD
2. FPD
3. NPD
4. CDP-4A data processor
5. N-2000 chromatogram workstation



Shipping Package


Dimension: 850x680x750mm
N. W.: 50kg
G. W.: 65kg


 
INESA is an experienced gas chromatograph manufacturer in China. Apart from chromatograph, we also provideatomic absorption spectrophotometer, spectrophotometer, flame photometer, multi-parametermeter, meltingpoint apparatus, pH meter and more.


Recommended Products
Hot Products