Home Medical Needle (Tubing) Tester

Medical Needle (Tubing) Tester

1. ZR9626-D Medical Needle (Tubing) Resistance Breakage Tester
The tester adopts a 5.7-inch color LCD to show menus: type of tubing wall, bending angle, designated metric size of the tubing, distance between rigid support and point of application of bending force, and number of bending cycles. PLC realizes program setup, which ensures tests are performed automatically.


Tubing wall: normal wall, thin wall, or extra thin wall is optional
Designated metric size of the tubing: 0.2mm~3.4mm
Frequency under test: 0.5Hz
Bending angle: 15°, 20° and 25°
Bending distance: with an accuracy of ±0.1mm
Number of cycles: to bend the tubing in one direction and then in the opposite direction, for 20 cycles


2. ZC15811-F Medical Needle Penetration Force Tester
The tester adopts a 5.7-inch color touch screen to show menus: nominal outside diameter of needle, type of tubing wall, test, test times, upstream, downstream, time and standardization. It displays maximum penetration force and five peak forces (i.e. F0, F1, F2, F3 and F4) in real time. And the built-in printer can print the test report.


Tubing wall: normal wall, thin wall, or extra thin wall is optional
Nominal outside diameter of needle: 0.2mm~1.6mm
Load capacity: 0~2N with an accuracy of ±0.01N
Movement speed: 100mm/min
Skin substitute: polyurethane foil compliant with GB 15811-2001
Way of printing: after each test finishes, all test data including the mean value and the deviation value of penetration forces can be printed.


3. ZG9626-F Medical Needle (Tubing) Stiffness Tester
The tester is controlled by PLC, and it adopts a 5.7-inch color touch screen to show menus: designated metric size of the tubing, type of tubing wall, span, bending force, maximum deflection, print setup, test, upstream, downstream, time and standardization. And the built-in printer can print the test report.


Tubing wall: normal wall, thin wall, or extra thin wall is optional
Designated metric size of the tubing: 0.2mm~3.4mm
Bending force: 5.5N~60N with an accuracy of ±0.1N
Load velocity: to apply downwards at a rate of 1 mm/min to the tubing the specified bending force
Span: 5mm~50mm (11 specifications) with an accuracy of ±0.1mm
Deflection test: 0~0.8mm with an accuracy of ±0.01mm


4. LD-2 Breaking Force and Connection Fastness Tester
The tester is designed and manufactured according to YY0321.1 "Single-use puncture set for local anesthesia" and YY0321.2 "Single-use needle for anesthesia". It can test the minimum forces required to break the catheter, the union of the catheter and catheter connector, the bond between the hub and needle tube, and the connection between the stylet and stylet cap.


Displayable force range: adjustable from 5N to 70N; resolution: 0.01N; error: within ±2% of reading
Test speed: 500mm/min, 50mm/min, or 5mm/min; error: within ±5%
Duration: 1s~60s; error: within ±1s; with LCD display


 



Hengyi is a major manufacturer of medical needle testers, based in China. Our company offers a wide range of products, including medical electrical safety testers, checkweighers, inspection machines, etc. Our products embody high quality and are competitively priced. We can supply the full manufacturing chain of powder and grain bottling assembling lines, automatic unscramblers, etc. right here in China. Our low manufacturing cost helps you save purchasing cost. If you would like to know more about our products, please turn to the corresponding pages for detailed description.

Medical Needle (Tubing) Tester

Recommended Products
Hot Products