Home F8 Multi-parameter Patient Monitor

F8 Multi-parameter Patient Monitor

F8 Multi-Parameter Patient Monitor

The F8 multi-parameter patient monitor with the standard configuration of 3-lead or 5-lead ECG/HR, RESP, TEMP, SpO2/PR, and NIBP.  Optional parameters include 2-IBP, ETCO2, and the depth of anesthesia monitoring module.  Optional accessories include VGA, printer, wall-mount, and a portable trolley.  The 12.1 inch color TFT LCD screen supports up to 11 different waveforms on a single screen and implements a multi-language interface that accommodates up to 14 languages. The robust display modes include large font, NIBP trend, trend table, full lead ECG, and Oxy CRG.  A fully featured alarm system includes visual on-display, integrated alarm light, and a personated voice reminding alarm.  The multi-parameter patient monitor is capable of producing a drug calculation and titration table, real time S-T segment analysis, pace detection, as well as Arrhythmia analysis.  With a massive storage that can accommodate 720 hours of patient data, 1000 NIBP records, and 500 individual alarm events, the monitor is suitable for extensive monitoring of adult, pediatric, and neonatal patients.  This unit is isolated against defibrillation and can achieve wired or wireless connection with a central station system.

Configurations

F8 Multi-parameter Patient Monitor

● Standard

○ Optional

- Unavailable

ECG

SpO2/PR

2- TEMP

NIBP

RESP

2-IBP

Nellcor SpO2

VGA

Printer

50mm

Wall-mount

/Trolley

Central Station

Touch Screen

ETCO2 Mainstream/

Side-stream/

Micro-stream

Multigas /

O2

F8



Physical Specifications


Dimension and weight

Dimension

340mm (W)*280mm(H)*130mm(D)

Weight

standard configuration: 4 kg

Operation environment

Power requirements

AC 90-300V, 50/60HZ, 60W

Power

90VA

Temperature

5-40℃

Humidity

30-93% non-condensing

Safety

Meet the requirements of IEC60601 series

Patient range

Adult, neonate and pediatric patients


 


Performance Specification of the F8 Multi-Parameter Patient Monitor


Display

12.1 inch (800 x 600 pixel) color TFT LCD rolling and refreshing waveform display

Display mode

Standard, large font, NIBP trend, trend table, full lead ECG, Oxy CRG.

Interface

Parameter cable interfaceAC power input socketNetwork interface External VGA interface for connection of alternate display

Battery

Internal rechargeable lead-acid cell, 12V/2.0AH

Alarm indicator

Visual on-display and integrated alarm light

Audible alarm and personated voice reminding alarm

Data storage

720 hours monitoring data10000NIBP records storage 500 alarm events.

Printer

Built-in 50mm thermal printer (option), 3 traces, print speed 25mm/s & 50mm/s print width:50mm

Net working

Support connection to central monitoring station through LAN (TCP/IP) or wireless connection

ECG

Leads type

3(option)/5-lead : I, II, III, aVR , aVL , aVF , V

Sweep speed

12.5mm /s, 25mm/s,50mm/s

Gain selection

×0.25 , ×0.5, ×1, ×2

ECG waveform

3 channels

Operation mode

Diagnostic, monitor, surgery

Heart rate range

15 -350bpm

Accuracy

1bpm or ±1% whichever is great

Alarm

ST/ arrhythmia

Real time S-T segment detection, pace detection and arrhythmia analysis

Protection

Against electrosurgical interference and defibrillation



Respiration (RESP)


Method

Thoracic impedance

Operation mode

Auto/manual

Measurement range

Adult: 0-120BrPM; neonatal /pediatric: 0-150BrPM

Apnea alarm

Yes

Alarm

Audible and visual alarm, personated voice reminding alarm and alarm events recallable



NIBP


Method

Automatic oscillometric

Operation mode

Manual/Automatic/STAT

Auto measure time

Adjustable

Measurement unit

mmHg / kPa selectable

Test interval

1-480 minutes

Measurement range

Adult: 0-300 mmHg ; Child: 0-240 mmHg ; Infant: 0-150 mmHg

Over-pressure protection

Double safety protection

Accuracy of blood pressure measurement

The mean error shall be less than±5mmHgthe standard deviation shall be less than 8mmHG

Resolution

1 mmHg

Accuracy

±3bpm(non-motion) ±5bpm(motion )


 


SpO2


Measurement range

0%-100%

Resolution

1%

Accuracy

±2% (70-100%, adult/pediatric, non-motion ) ±3% (70-100%, neonate, non-motion ) ±3% (70-100%, adult/pediatric/neonate, motion ) 0-69% unspecified

Alarm range

1-100%

Pulse rate

Pulse range:

0-350bpm



Temperature (TEMP)


TEMP range

0-50℃

Resolution

0.1 ℃

Accuracy

±0.1℃ (excluding the sensor)

Channel

Dual-channel. provide T1;T2; △ T



IBP


Measurement range

-50-360mmHg

Channel

2 channels

Pressure transducer

Sensitive :5u V/V mmHg Impedance range :300-3000 Ω

Pressure names

ART ,PA ,CVP,RAP,LAP,ICP,P1,P2

Alarm range

-10-300 mmHg

PR from IBP

Range & Accuracy:

25-200bpm ±1bpm or ±1bpm which is greater 201-350bpm ±2%

Resolution

1bpm



Depth of Anesthesia


EEG sensitivity

/- 475μV

Noise

< 2uVp-p (1-150 Hz)

CMRR

> 100 dB

Input impedance

> 50mohm

Sample rate

1024 samples/s, 16 bits

Index and display update

IoC 0-99, filter 1-47 Hz, 1 sec. display update

EEG waveform

± 500 µV scaleable, filter 1–30 Hz, 3 sec. window length.

ESR

0-100%

EMG 0-100 arbitrary unit

Curve and bar

Index trend

5 min and 60 min

Digital output

Wireless BT interface (RFCOMM). 10 meters range

Alarms

IoC high and low user selectable limits

Artifact rejection

Automatic: user can disable.

Sensor impedance range

0 - 14kohm



EtCO2 (Main-stream / Side-stream / Micro-stream CO2)


Measurement range

0-150mmHg

Accuracy

±2mmHg(0-40mmHg)±5% of reading(41-76mmHg)±10% of reading (77-99mmHg)

Resolution

1mmHg

Sampling rate

70m 1/min 100m1/min

Sampling rate accuracy

0.15 or 15m1/min

Start-up time

30 Sec. the module enters the warming up status after the start up.1 minute later, it enters the ready-to-measure status.

Respiration rate

3-180 rpm

Respiration rate accuracy

±2BrPM (0-70BrPM) ±5BrPM (>70BrPM)

Response time

When measured with a neonatal water trap and a 2.5 m-long neonatal sampling line: < 3.5 s @ 100 ml/min <4 s @ 70 ml/min When measured with an adult water trap and a 2.5 m-long adult sampling line: <5.5 s @ 100 ml/min <7 s @ 70 ml/min

Delay time

When measured with a neonatal water trap and a 2.5m-long neonatal sampling line: <3 s @ 100 ml/min <3.5 s @ 70 ml/minWhen measured with an adult water trap and a 2.5m-long adult sampling line: <5 s @ 100 ml/min <6.5 s @ 70 ml/min



Multi-Gas/O2


Method

Infrared absorption

Gas sorts

CO2, N2O, Des, Iso, Enf, Sev, Hal, O2,(Optional paramagnetic sensor)

Measurement range

CO2: 0~30% N2O : 0~105% O2, :0~105%Enf , Iso , Hal: 0~30% Sev : 0~30%Des:0~30%

Data output

Fi and ET values

Respiration rate

±1BrPM(2~60BrPM)61~100BrPM (unspecified)

Other

Up to 4 waveforms displayedAgent mixture detection MAC value displayed


 