Home F6 Multi-parameter Patient Monitor

F6 Multi-parameter Patient Monitor

256mm (W) x 209mm (H) x 185mm(D)
Weight:     4 Kgs (with battery)
Power Supply : 100~240V;   AC: 47~ 63 Hz
12 VDC
Alarm Indication: Visual and audible alarm

ECG
Leads:  I, II, III, aVR,  aVL,  aVF, Cx
Gains: 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, auto
Input impedance: ≥5MΩ (d.c.~60Hz)
Common Mode Rejection Ratio: ≥85dB(diagnostic mode)
≥105dB(monitoring & surgica mode)
Bandwith: 0.5-40Hz (monitoring mode)
         0.05-120Hz(diagnostic mode)
         1-25Hz (surgical mode)
Protection: Segregation breakdown voltage 4000VAC, 50/60Hz
Resistant to high-frequency electrotome, myoelectricity, AC, drifting interference
Indication of Detachment of electrodes: Available
Playback of Cardiogram: Available
Scanning Speed: 12.5mm/s, 25mm/s, 50mm/s
Heart Rate Range:10-300BPM
Accuracy:  ±1BPM

NIBP
Method: Automatic oscillometric
Operation Modes: Singles (manual), Cycle (automatic), Rapid (continuous)
Measurement : Systolic pressure/diastolic pressure/mean
Range: Adult: 10-270mmHg
Pediatric: 10-235mmHg
Neonate: 10-135mmHg
Unit: mmHg / Kpa selectable
Auto Measurement: 1,2,3,4,5,10,15,20,30,45,60,90 minutes, 2,4,8,12 hours
600 groups of NIBP measuring data review

RESP
Method: Impedance Measurement
Measuring Range: 0-120bpm
Alarm Limit: 0-120bpm, upper and lower limit continuously adjustable
Apnea alarm: Available
 
Digital SPO2 (Pulse Oximetry)
Range: 0-100%
Accuracy:  ±2% (70%~100%); 
±3%(35%~ 69%)
Unspecified (0~34%)
Range & Accuracy of Pulse Rate : 25~250BPM ± 1BPM
 
TEMP
Channel : 2 channels
Range : 0~50℃
Unit : Celsius, Fahrenheit
Resolution : 1℃
Accuracy : ±0.1℃ (25℃~50℃)
 
Environment Conditions:
Temperature:
Operation 0℃-50℃
Storage / Transportation : -40℃-70℃
Humidity:
Operation: 10%~95%,  noncondensing
Storage / Transportation : 0%~93% noncondensing
Atmospheric Pressure:
Operation: 700~1060hPa, 
Storage / Transportation : 500~1060hPa
 


Recommended Products