Applications:
Classic-90 is design for simple and accurate vital signs assessment or continuous monitoring, this monitor is the definition of "advanced monitoring made affordable."Features:
·Parameters: SpO2, NIBP, Pulse, TEMP·Network capability with central monitoring system
Specifications:
ECG | Heart rate display range | 15bpm ~ 350bpm | |
Heart rate display accuracy | ± 1% or ± 2bpm, whichever is greater | ||
Sweeping speed | 12.5mm/s, 25mm/s, 50mm/s | ||
Sensitivity selection | ×1/2, 5mm/mV ×1, 10mm/mV ×2, 20mm/mV | ||
CMRR | ≥ 89dB | ||
RESP | Measuring range | 0rpm~120rpm | |
Accuracy | ±5% or ±2 rpm | ||
Alarm limits setting range | 0rpm~120rpm | ||
TEMP | Measuring range | 25.0℃~45.0℃ | |
Measuring accuracy | ±0.2℃ | ||
Responding time | ≤150s | ||
NIBP | Measuring method | Oscillometric Technique | |
Measuring range | 0 mmHg~300mmHg | ||
Accuracy | Maximal mean difference | ±5 mmHg | |
Maximal standard deviation | 8 mmHg | ||
Measurement range | Adult | SYS 40~255 mmHg | |
MAP 20~215 mmHg | |||
DIA 10~195 mmHg | |||
Infant | SYS 40~200 mmHg | ||
MAP 20~165 mmHg | |||
DIA 10~150 mmHg | |||
Neonate | SYS 40~135 mmHg | ||
MAP 20~110 mmHg | |||
DIA 10~95 mmHg | |||
SpO2 | Measuring range | 35%~100% | |
Low perfusion capability | 0.4%~5% | ||
Accuracy | Not greater than 3% for SpO2 range from 70% to 100% | ||
PR | Measuring range | 30bpm~240bpm | |
Measuring accuracy | ±2bpm or ±2%, whichever is greater |