N2/O2 Analyzer
P860 series analyzer isthe online intelligent measurement instrument adopting advanced ion currentoxygen sensor with single chip technology. N2% = 100%-O2%.The N2/O2 analyzer hasbeen widely used in the fields of nitrogen separation from air, fertilizer,petrochemicals and biological fermentation.
Features of the N2/O2 Analyzer
1. New patternlong-life conversion ion current tester with fast response and high sensitivity
2. Sensor workingstate automatically checking and unique double-line working change modeextending the sensor’s life.
3. The N2/O2 analyzer has N2 or O2 concentration display
4. Setting alarmvalue freely
5. High precision,automatic temperature compensation, eliminating the impact of environmentaltemperature
Technical Specifications of the N2/O2 Analyzer
Monitor | Five-digital dynamic LED monitor, visible display window 90mm×25mm |
Measure Range | 10~2000ppm, 0~2%, 0~21% O2 |
Measure Mode | Air inflowing or convection diffusing |
Sample Gas Flow | 150~300ml/mim |
Accuracy | ±5% in 10~2000ppm O2 range ±2% in 0~2%O2 range ±1% in 0~21%O2 range |
Response Time | 90% of FS reading ≤20 seconds |
Ambient Temp. | -10℃ to 50℃ |
Ambient Humidity | <80%RH |
Stability | <±1%.FS (168h) |
Signal Output | 4~20mA |
Power Supply | 220±10% VAC, 50/60Hz, the power loss ≤6VA |
Sensor Life | 20,000 hours at 25℃ and 1 atm |
Warranty | 12 months analyzer; 12 months sensor |
Weight | 1 Kg |
Appearance Dimension | 160 mm × 80 mm × 176 mm (W×H×D) |
Opening Size | 151 mm × 75 mm |
ModelChoice for theN2/O2 Analyzer
Item | Model | Measure range |
1 | P860-3O | 21.0%~1000ppm O2 |
2 | P860-4O | 21.00%~100ppm O2 |
3 | P860-5O | 21.000%~10ppm O2 |
4 | P860-3N | 79.0%~99.9% N2 |
5 | P860-4N | 79.00%~99.99% N2 |
6 | P860-5N | 79.000%~99.999% N2 |
CHANGAI is a professional N2/O2 analyzer manufacturer in China. We also offerflue gas oxygen analyzer, high purity portable oxygen analyzer, aluminumenclosure and more.